Instructions for Use Fabius MRI SW 3.n
87
Operation
During N
2
O failure O
2
may still be administered. No
alarm sounds. The float in the N
2
O measuring tube
drops to zero.
After each O
2
supply failure the O
2
shortage alarm
must be reactivated.
However, S-ORC is not an oxygen-specific moni-
toring device and therefore cannot provide any pro-
tection against the effects of accidental use of the
wrong gas.
Therefore
always
monitor the O
2
concentration.
CAUTION
Risk of inaccurate measuring values
After the O
2
supply has been restored, a supply
pressure of at least 2.7 kPa x 100 must be applied
for at least 20 seconds before another O
2
short-
age signal can be emitted.
During this period, do not activate any devices
that consume O
2
(e.g. O
2
flush, O
2
fresh gas flow
or secretion aspiration).
WARNING
Damage to materials and health risks
Never switch off fresh-gas flow before the
vaporizer is switched off. A vaporizer must
never be left switched on without a fresh-gas
flow, because high-concentration anesthetic
vapor may leak into machine lines and ambi-
ent air, causing damage to materials and
health risks.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...