Instructions for Use Fabius MRI SW 3.n
Calibrate Flow Sensor
z
Press the »
Calibrate Flow Sensor
« key on the
Standby screen. The Standby soft key window
is replaced by instructions for performing the
calibration.
z
Follow the instructions on the screen:
– Close all fresh gas control valves.
– Remove expiratory hose from breathing sys-
tem.
– To start Flow Sensor Calibration, press rotary
knob.
z
When the calibration begins, the instructions
are removed, and a »
Flow Calibration in
progress
« message is displayed above the
Standby soft keys.
z
When the calibration is completed, one of two
messages is posted above the Standby soft
keys: »
Flow Calibration completed - recon-
nect expiratory hose
« or »
Flow Calibration
Failed
«.
Troubleshooting Flow Calibration Failure
If the flow sensor cannot be calibrated, retry the cal-
ibration.
If the flow sensor still cannot be calibrated, call
DrägerService or your local authorized service
organization.
599
52
5
526
527
52
8
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...