Instructions for Use Fabius MRI SW 3.n
155
Configuration
1
During the calibration period, the word “
CAL
”
replaces the O
2
value in the oxygen monitoring
window. Calibration time is approximately
15 seconds. Upon successful completion of the
calibration, the O
2
measurement value will be
restored.
If the O
2
sensor cannot be calibrated, replace the
O
2
capsule in the O
2
sensor housing (see
page 189).
If the O
2
sensor still cannot be calibrated, call
DrägerService or your local authorized service
organization.
Des Comp On/Off
CAUTION
The Des Comp On/Off function is not relevant for
MRI use.
CAUTION
The Des Comp On/Off parameter remains in the
Off position on the Fabius MRI.
WARNING
The Desflurane vaporizer is not useable in an
MR environment. In an MR environment func-
tionality of the Desflurane vaporizer will be
compromised.
515
1
503
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...