222
Instructions for Use Fabius MRI SW 3.n
Anesthetic Gas Scavenging System
P
Check the hose connections.
P
Adjust the flow regulator to place the float
between the "Minimum" and "Maximum"
marks.
P
Change to Man/Spont mode.
Press and hold the O
2
flush button and ver-
ify that airway pressure is <10 cmH
2
O
(hPa) with Y-piece occluded.
Wait until flow curve is refreshed.
P
Close all flow control valves on the
machine, with Y-piece occluded, and verify
that airway pressure is
≥
0.5 cmH
2
O (hPa)
Manual Ventilation Bag for Emergency Ventila-
tion
Check that the bag is functioning correctly
by pumping manually.
When the bag is squeezed, air must audi-
bly and tangibly flow out of the mask cone;
when the bag is released, it must rapidly
recover its original shape.
Block off the mask connector (cone) with
the ball of your thumb: you should only be
able to squeeze the bag a little.
P
Before Connecting to Patient
Verify that
– all vaporizers are off (the handwheels
are set to zero),
– the APL Valve is set as desired,
– all flowmeters indicate 0,
– the patient suction is level adequate,
and
– the breathing system is ready to use
(the bag is in place and all hoses are
connected properly
If any check can not be carried out satisfacto-
rily, the machine must not be used.
Daily Checkout Signature
Name
Date
Pre-use Checkout Signature
Name
Date
Pre-use Checkout Signature
Name
Date
Pre-use Checkout Signature
Name
Date
Pre-use Checkout Signature
Name
Date
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...