9
Adding Modules
Adding an Analog Output Interface
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
491
Adding an Analog Output Interface
This section provides instructions for installing the Analog Output Interface (AOI) on your
TRACE 1300/1310 GC.
To install an analog output interface
1. Put the GC in standby condition.
2. Cool the oven, injectors and detectors to room temperature.
3. Power off the GC.
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
4. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
1. Remove the cover of the external modules housing where installing the Analog Output
Interface module. See
Figure 524.
Housing Cover Removal
a.
Using a T20 Torxhead screwdriver, unscrew and remove the left or right housing
cover screws, then remove the cover from the housing.
2. Install the Analog Output Interface into the housing.
a.
Loosen the two hexagonal screws under the module. See
.
WARNING
This operation must be carried out by authorized and trained Thermo Fisher
Scientific Service Field Engineers.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.
Left/Right Housing Covers
Left Cover
Screws
Right Cover
Screws