11
Upgrade Equipment
Upgrading a Stand Alone TRACE 1300/TRACE 1310 to MS Version
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
565
8. Make the electrical connections. See
Figure 622.
Electrical Connections
a.
Connect the heater cable coming from the transfer line to the connector marked
Heater 1
or
Heater 2
on the front of the module
b. Using the cable provided, connect the 15-pin female connector marked
GC Bus
on
the module to a
Bus
interface on the back of the GC.
c.
Plug the power cable to the AC Input connector on the front of the module and to
the wall outlet. The LED marked
On
lights.
9. Close the front door of the GC.
Tip
To avoid the contact with the hot air from the vents, it is suggested to gather
the electrical cables into the cables holder.
Aux Temperature/Cryo Module
Transfer Line
Cables Holder