8
Installing Optional Kits
Installing the Hot Injection Adapter Kit on the SSL/SSLBKF Injector
402
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
a.
Open the front door of the GC.
b. Loosen the retaining nut from the injector fitting on the upper interior wall of the
GC oven.
c.
Remove the analytical column with its nut and ferrule from the bottom of the
injector.
7. Put the autosampler away if present.
8. Remove the SSL/SSLBKF injector module from its seat.
a.
Open the module flap cover.
b. Using a T20 Torxhead screwdriver, unscrew the three captive fixing screws.
c.
Throw upward the module from its seat of the injector housing. Place the
SSL/SSLBKF module on a clean surface.
9. Replace the module flap cover with the one provided. See
.
Figure 410.
Module Flap Cover for Hot Injection
10. Reinstall the injector module into the main frame.
a.
Open the module flap cover.
WARNING
Make sure the o-ring remains into its seat on the gas connection.
Do not install the module if the o-ring is missing
.
O-ring
Note
The fixing holes present on the top of the module flap cover can be used for
making more stable a sampling device that requires to be seat directly on the injector.
Module Flap Cover
Fixing Holes