5
GC Main Frame Advanced Maintenance
Removing/Replacing the GC Right Side Panel
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
199
b. Slide the panel towards the back of the instrument up to the stop.
c.
Remove the panel pulling it outwards being aware that the ground wire is attached to
the panel. See
Figure 185.
Left Panel Removal
d. Unplug the ground wire from the panel.
6. Replace the panel.
a.
Plug the ground wire to the panel.
b. Reinstall the panel proceeding in the reverse order in which it was removed.
Removing/Replacing the GC Right Side Panel
To remove the right side panel with the side panel molded of the GC
1. Put the GC in standby condition.
2. Cool the oven, injectors and detectors to room temperature.
3. Power off the GC.
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
Note
Pay attention to the positioning of the ground wire plug, so it can be
reconnected in the same way it was removed.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.