3
Performing Injectors Routine Maintenance
Maintaining an Instant Connect Helium Saver Injector (HeS-S/SL)
110
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
• Replacing the active carbon filters on the carrier gas line and split line
The active carbon filters must be replaced depending on the volume of solvent injected in
the time.
See
“Replacing the Carrier and Split Lines Filters”
.
• Replacing the body head o-rings
The internal (carrier line) and external (purge line) o-rings of the body head must be
replaced when there are leaks present.
See
“Replacing the Body Head O-Rings”
.
Before maintaining the injector, read the following warning:
WARNING
The injector fittings could be hot. Carry out all the operation at low
temperature to avoid burns. Therefore, before beginning the sequence, the injector must
be cooled to room temperature.
CAUTION
When handling organic solvents, you must take precautions to avoid health
hazards.
Materials needed to maintain a HeS-S/SL
Septum
Tweezers
Glass liner
Liner seal (O-ring)
Ultrasonic cleaner
Mixture 1:1 methanol/acetone
Carrier gas line and/or split gas line active carbon filters
Body head internal o-ring
Body head external o-ring
T20 screwdriver
Note
For maintaining or replacing any other component not listed in this section, see
Chapter 6, “Injectors Advanced Maintenance.”
•
“Baking-out Contaminants from SSL, SSLBKF, HeS-S/SL, PTV, and PTVBKF
Injectors”
•