Performing Injectors Routine Maintenance
Maintaining an Instant Connect Helium Saver Injector (HeS-S/SL)
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
115
5. Put the autosampler away if present.
6. Open the module flap cover.
7. Replace the filter.
a.
Remove the filter to replace from its seat by turning it counter-clockwise.
b. Install the new filter, with o-rings, in its seat by turning it clockwise.
8. Close the module flap cover.
9. If present, move the autosampler towards the module to restore the original alignment.
10. Wait two or three minutes for helium to purge the inlet.
11. Turn the inlet flow back on and enable the inlet heater.
12. Set the normal injector, detector, and GC working conditions.
Replacing the Body Head O-Rings
To replace the body head o-rings
Figure 101.
HeS-S/SL Injector: Body Head O-Rings Replacement
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
Body Head Internal O-ring (Carrier gas Line)
Body Head External O-ring (Purge gas Line)