9
Adding Modules
Adding a FID, TCD/TCD In-Series, ECD, or FPD Detector Module
432
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
6. Remove the dummy module from the position where the detector module will be
installed.
a.
Open the module flap cover.
b. Using a T20 Torxhead screwdriver, unscrew and remove the two captive fixing
screws.
c.
Keeping the dummy module flap cover open, lift up the module from its seat in the
injector/detector housing. Place the dummy module on a clean surface.
d. Remove the gas block plug from the gas connections by unscrewing its fixing screw
using a T20 Torxhead screwdriver. See
.
Figure 447.
Detector Gas Block Plug
7. Open the front door of the GC.
8. Plug the detector module into the main frame.
a.
Open the module flap cover.
ATTENTION
The FPD detector must be installed in the
back
site of the detector housing.
For details about the installation of the FPD detector module, see the section
.
Gas Connections Without Gas Block Plug
Gas Block Plug on Detector Seat
O-rings
WARNING
Make sure all the four o-rings are placed into their seats on the gas
connection. See
.
Do not install the module if the o-rings are missing.