1
Installation
Connecting the PTV/PTVBKF Cryogenic System
18
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
3. Connect the cryo valve to the
Aux Temperature/Cryo Module
.
a.
Connect the cryo solenoid valve to the 2-pin connector marked
Cryo Valves - Oven
using the cable provided. See
Figure 18.
Cryo Valves: Oven
b. Using the cable provided, connect the 15-pin female connector marked
GC Bus
on
the module to a
Bus
interface on the back of the GC.
c.
Plug the power cable to the AC Input connector on the front of the module and to
the wall outlet. The LED marked
On
lights up after the GC power on.
Connecting the PTV/PTVBKF Cryogenic System
If the PTV or PTVBKF injector of your TRACE 1300/TRACE 1310 is equipped with a
cryogenic cooling option, you need a supply of coolant, carbon dioxide (CO
2
) or liquid
nitrogen (LN
2
).
PTV/PTVBKF Cryo System with Carbon Dioxide Connection
This cryo system consists of a tube for Carbon Dioxide inserted into the PTV/PTVBKF
module, and fitted with a 1/8-in. solenoid valve for carbon dioxide mounted on a bracket
fixed on the back of the GC. See
.
IMPORTANT
The Auxiliary Temperature/Cryo must be powered at the same line voltage
of the main GC system.
Note
For further details regarding the installation of the Aux Temperature/Cryo module,
“Adding an Aux Temperature/Cryo Module”
Oven Cryo Valve