9
Adding Modules
Adding a GSV Injector Module
426
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
6. Put the autosampler away if present.
7. Remove the dummy module from the position where the injector module will be
installed.
a.
Open the module flap cover.
b. Using a T20 Torxhead screwdriver, unscrew the three captive fixing screws.
c.
Keeping the dummy module flap cover open, lift up the module from its seat into the
injector housing. Place the dummy module on a clean surface.
d. Remove the gas block plug from the gas connections by unscrewing its fixing screw
using a T20 Torxhead screwdriver. See
.
Figure 441.
Injector Gas Block Plug
8. Open the front door of the GC.
9. Plug the Gas Sampling Valve module into the main frame.
Gas Connection Without Gas Block Plug
Gas Block Plug on Injector Seat
O-ring
WARNING
Make sure the o-ring is placed into its seat on the gas connection.
.
Do not install the module if the o-ring is missing.