10
Adding Systems
Adding the Oven Cryo System
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
509
12. Mount the left side panel proceeding in the reverse order in which the left side panel was
removed.
13. Open the gas supplies.
14. If other external modules are present, plug the power cable to the AC Input connector of
each external module, and to the wall outlet.
15. Power on the GC.
a.
Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker), located at the back of the instrument, to the
position I.
16. Configure and enable the cryogenic system through the user interface of your GC, or
through the CDS in use. Refer to the
TRACE 1300 and TRACE 1310 User Guide
.
a.
Configuration and enabling through the touch screen
.
i.
In the main menu select the
Configuration
icon. The Configuration menu
appears.
ii. In the Configuration menu, select the
Oven
icon to open the relevant submenu.
iii. Set the
Cryogenic
parameters.
–
Cryogenic Type
— Enable or disable the cryogenic system when it is
installed and configured with Carbon Dioxide or Liquid Nitrogen as a
coolant. Select between LN
2
, CO
2
, or none.
–
Cryo timeout
— Enter the time at which the cryo system will be disabled.
Enter a value from 0–30 min.
–
Cryo Threshold
— Specify the temperature at which the cryo system begins
to supply the coolant. Enter a value from 40-200 °C.
iv. Return to main menu.
b.
Configuration and enabling through the Chromatography Data System
.
i.
In the
Configuration
window select the
Auxiliary
tab.
ii. Select the
Auxiliary control module
check box to enable the setting for the
auxiliary control of the module option installed on your GC.
iii. Select the
Oven cryogenics
check box.
iv. Select the
Cryo type
used by your cryogenic option. Choose one: Liquid
Nitrogen or Carbon Dioxide.
v.
Open the
Oven
page. Select the
Cryogenics enable
check box to enable the
cryogenic system.
vi. In
Cryo threshold
text box specify the temperature at which the cryo system
begins to supply the coolant. Enter a value from 40-200 °C.