1
Installation
Column Installation Requirements
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
51
•
Ferrules
— Graphite ferrules, graphite Vespel® ferrules, and encapsulated graphite
ferrules are used for many column connections. See
Table 4.
Column Retaining Nuts for Injectors and Detectors
Retaining Nut Type
Injector/Tee Connector
Detector
1/4” Hexagonal
retaining nut
SSL, SSLBKF
FID, NPD, TCD,
ECD, FPD
M4 Split retaining
nut
PTV, PTVBKF,
SSLBKF
(Tee Connector)
Table 5.
Ferrules for Injectors and Detectors
Ferrule Type
Injector/Tee
Connector
Detector
Temperature
Limit
Graphite ferrule
SSL, SSLBKF
FID, NPD, TCD,
ECD, FPD
450 °C
Graphite Vespel®
ferrule
SSL, SSLBKF
FID, NPD, TCD,
ECD, FPD
350 °C
Encapsulated
graphite ferrule
(Graphpack)
See the
Note
below.
PTV, PTVBKF,
SSLBKF
(Tee Connector)
450 °C
Graphite ferrules
are a soft material that is porous to oxygen, making them suitable for most
applications except GC/MS interface connections. These easy-to-use ferrules form a soft grip with the
column and provide a stable seal. Graphite ferrule are re-usable.
Graphite/Vespel®
ferrules have a long lifetime and are compatible with GC/MS. These ferrules form
a strong grip with the column and cannot be reused as they form a permanent seal with the column.
They have a temperature limit of 350 °C, but must be re-tightened after initial temperature cycles.