Column Installation Requirements
50
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Configuring the Data System
You should use the self adhesive labels found in the standard outfit of the Data System to
annotate the IP address and the TCP Port that are set in the instrument. Stick the label where
it can be easily found when necessary to configure the data system.
1. Launch the Data System. In the relevant Configuration page specify a direct TCP/IP
address.
2. Enter the IP address of the GC and set the socket used.
Standard operation parameters are set by default; however, your GC may be hidden
behind a Firewall that may prevent Port 2551 from being used.
You can set an alternative port number. The number of the socket entered in this box
must correspond to the port assigned to the instrument setup.
Column Installation Requirements
The column is where the separation takes place. It should be positioned inside the oven on the
column rack. The column ends should align correctly with the injector and the detector
bottom fittings. The GC oven controller accurately controls the column temperature.
Each column has a maximum recommended operating temperature. To protect the column
from excessively high temperatures, remember to set the
Maximum temperature
parameter
for the oven.
Related Topics
•
•
“Installing the Adapters for Encapsulated Graphite Ferrules”
•
Using the Correct Fittings
To connect a capillary column to the injector and detector base body, you must use the proper
column ferrules and retaining nuts.
•
Retaining Nut
— Dedicated retaining nuts are required to connect capillary columns to
injector and detector base bodies, and the
Tee
connector. See
IMPORTANT
If the connection is performed through hubs over a 10 Mbit/s network, it is
suggested that no more than five GC units be connected on the same network trunk.
If you have a switched network, this warning does not apply.
Note
For detailed information on maximum operating temperature, refer to the
column
manufacturer’s instructions
provided with the column.