Detectors Advanced Maintenance
Measuring the NPD Gas Flows
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
283
Figure 264.
Reinstall Thermionic Source (2)
19. Reconnect the source assembly cable to the NPD cable and twist the ring to lock the
connection. See
.
Figure 265.
Reinstall Thermionic Source (3)
20. Close the module flap cover.
21. Plug in the signal cable into its contact on the detector module. See
Figure 266.
Plug NPD Signal Cable
Collecting Electrode Cable
Collecting Electrode Straight Jack Connector