10
Adding Systems
Adding an Auxiliary Gas System
526
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
c.
Slide the panel towards the back of the instrument up to the stop.
d. Remove the panel pulling it outward being aware that the ground wire is attached to
the panel. See
e.
Unplug the ground wire from the panel.
Figure 569.
Left Side Panel Removal
3. Prepare the duct for the installation of the auxiliary gas interface.
a.
Remove the partial cut shaped plate from the exterior wall of the oven box for
accessing the insulating material. See
.
Figure 570.
Perform the Duct for the Auxiliary Gas Interface (1)
Partial Cut Shaped Plate