Performing Detectors Routine Maintenance
Maintaining a Nitrogen Phosphorous Detector (NPD)
152
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
10. Unscrew and pull out the straight jack connector of the collecting electrode cable from
the collecting electrode bulkhead jack. See
.
Figure 128.
Collecting Electrode Cable Removal
11. Remove the thermionic source.
a.
Using a T6 Torxhead key, loosen the dowel which fix the thermionic source
connector to the thermionic source support. See
.
Figure 129.
NPD Thermionic Source Removal (1)
Collecting Electrode Cable
Collecting Electrode Straight Jack
Connector
Thermionic Source Connector
Dowel
Thermionic Source Support