8
Installing Optional Kits
Connecting a GSV Backflush System
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
373
b. Cut 2–3 cm from the column end.
c.
Insert the column into the detector base body and slide the ferrule up to the detector
base.
d. Finger-tighten the column retaining nut until it starts to grip the column.
e.
Push the column through the detector base according to the detector in use.
f.
Use the 6 mm wrench to tighten the retaining nut. Use enough pressure necessary to
obtain a good seal without overtighten.
To install of a second backflush system
If a second SSL/PTV Backflush system is installed, you can connect both the systems by using
two Tee connectors mounted on the same mounting bracket as schematically shown in
Figure 370.
Schematically Connection of a Second Backflush System
Connecting a GSV Backflush System
This sections provides instructions for connecting your Gas Sampling Valve module with
precolumn and analytical capillary column into the oven of the TRACE 1300/1310 GC.
Refer to the chapters Gas Sampling Valve (GSV) Module in the
TRACE 1300/TRACE 1310
User Guide
.
The Gas Sampling Valve backflush system is provided with the kit PN 19050764.
Tip
Turn the Tee connector as required to allow the installation of the two precolumns.
Capillary Column
to Back Detector
Precolumn
to Back Injector
Capillary Column
to Front Detector
Precolumn
to Front Injector
BackFlush Line
Backflush Line