7
Detectors Advanced Maintenance
Cleaning or Replacing the FPD Filter-side Heat Shields
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
303
8. Remove the filter-side heat shield.
a.
Loosen the three Allen screws that fix the filter support to the flange of the detector
body and remove it. See
Figure 290.
Filter-side Heat Shield Removal (1)
b. Turn down slowly the detector body and let the filter-side heat shield come out from
the flange. Pay attention not to lose the o-ring inserted in the filter support.
Figure 291.
Filter-side Heat Shield Removal (2)
9. Insert a new filter-side heat shield.
CAUTION
If the filter installed is equipped with the spacer (see the kits PN 19050785 and
PN 19050786), first remove the spacer, and then the filter.
Interferential Filter
Spacer
Flange
Filter Support
Filter-Side Heat Shield
Viton
®
O-Ring