8
Installing Optional Kits
Installing the Large Volume Splitless Kit
354
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Installing the Large Volume Splitless Kit
This section provides instruction for installing the Large Volume Splitless kit (PN 19050725)
on your TRACE 1300/TRACE 1310.
The Large Volume Splitless injector is a setup of the standard splitless injector, where the
introduction of large amount of liquid samples can be performed manually, or with the
TriPlus RSH, TriPlus 100 Liquid Sampler, or AI/AS 1310 autosampler.
Large Volume-Splitless kit includes:
• Two dummy filters
• Two o-rings for dummy filters
• An deactivated connector (press-fit)
• An uncoated precolumn (5 m x 0.32 mm ID)
• A dedicated splitless liner (set of 5)
• LV-SL Assistant software
To install the large volume splitless kit
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
3. Turn the carrier gas off, and wait for the carrier pressure to go to zero.
4. Put the autosampler away if present.
5. Open the module flap cover
6. Replace the filters with the dummy filters.
a.
Remove both the filters from their seats by turning them counter-clockwise.
Содержание TRACE 1300