Installing Optional Kits
Installing the Oven Exhaust Kit
326
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Installing the Oven Exhaust Kit
The oven vents at the back of the GC discharge hot air up to 450 °C (842 °F) during cooling.
Installing the optional Oven Exhaust Kit (PN 19050760), the hot air from the oven vents can
be carried away to a fume hood or other exhausting devices
.
The kit includes two air ducts and two 3-m length extensible aluminium tubes.
See
Figure 325.
Air Duct and Extensible Aluminium Tube
To install the oven exhaust kit
1. Put the GC in standby condition.
2. Cool the oven to room temperature.
3. Place each air duct on the oven vent inserting the hooks of the air duct into the relevant
vertical slots provided on the GC back panel.
4. Place an end of each extensible aluminium tube on each air duct.
5. Connect the other end of the extensible aluminium tube to an exhausting device.
6. Set the normal GC working conditions.
Air Duct
Extensible Aluminium Tube
Air Duct
Extensible Aluminium Tube
Slot