8
Installing Optional Kits
Installing the HS Adapter Kit on the SSL/SSLBKF Injector
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
345
b. Loosen the retaining nut from the injector fitting on the upper interior wall of the
GC oven.
c.
Remove the analytical column with its nut and ferrule from the bottom of the
injector.
6. Remove the SSL/SSLBKF injector module from its seat.
a.
Open the module flap cover.
b. Using a T20 Torxhead screwdriver, unscrew the three captive fixing screws.
c.
Lift the module from its seat in the injector housing. Place the SSL/SSLBKF module
on a clean surface.
7. Replace the module flap cover with the one’s provided. See
Figure 346.
Module Flap Cover for the HS Adapter
8. Reinstall the injector module into the main frame.
a.
Open the module flap cover.
b. Place the injector module in its seat. Be sure to insert the 25-pin male connector, on
the bottom of the module, into the 25-pin female connector on the injector seat of
the injector housing.
c.
Use a T20 Torxhead screwdriver to tighten the three captive fixing screws without
overtightening.
WARNING
Make sure the o-ring remains into its seat on the gas connection.
Do not install the module if the o-ring is missing.
O-ring
Module flap cover