9
Adding Modules
Adding a Helium Saver Injector Module
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
457
The composition of the gas mixture in the inlet will be mostly helium and allow good
sensitivity. For bulk nitrogen leak detection, the GC column exit can be plugged, and the
automated pressure drop leak detection of the GC used for determining the pressure drop.
The helium delivery block should be parked in the rear “blocked” position of
leak checking is done based on pressure drop. The block must be replaced in the front
position to resume operation.
Figure 475.
Removing the Helium Tube from the Gas Delivery Block