8
Installing Optional Kits
Connecting a SSL/PTV Backflush System
360
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Connecting a SSL/PTV Backflush System
This section provides instructions for connecting your SSLBKF/PTVBKF injector modules
with precolumn and analytical capillary column into the oven. Refer to the chapters SSLBKF
Injector Module and PTVBKF Injector Module in the
TRACE 1300/TRACE 1310 User
Guide
. This system operates up to 300 °C.
The operation consists of the following operating procedures:
•
“To assemble the Tee connector on the mounting bracket”
•
“To install the mounting bracket”
•
“To connect backflush line, precolumn, and capillary column”
•
“To install of a second backflush system”
To assemble the Tee connector on the mounting bracket
1. Remove the nut and the top end of the mounting bracket. This nut secures the two
sections of the mounting bracket. See
.
CAUTION
Before starting make sure the SSLBKF/PTVBKF injector module is correctly
installed into its seat.
Bottom
Tee Connector Assembly
Nut
Top
Mounting Bracket
Retaining Nut
Retaining Nut
Graphite/Vespel Ferrule
Graphite/Vespel Ferrule
Backflush Line Fittings
Nut
Clip
Clamp
Tee Connector
Threaded Shaft