8
Installing Optional Kits
Installing the HS Adapter Kit on the SSL/SSLBKF Injector
346
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
9. Remove the top parts of the injector. See
.
Figure 347.
Injector Top Parts Removal
a.
Unscrew the septum cap of the injector.
b. Unscrew the ring nut.
c.
Remove the septum holder/liner cap with septum from the injector body head.
10. Replace the liner.
a.
Use tweezers to remove the current liner with the liner seal (o-ring) from the injector.
b. Holding the HS/SPME liner with tweezers, place a new liner seal over the liner.
c.
Insert the liner into the injector and push it gently towards the bottom of the injector.
11. Install the HS adapter on the top of the injector. See
CAUTION
To maintain the correct alignment the screws must be tightened in turn.
Tighten each screw only a small amount before moving to the next screw. Repeat until all
are secure.
IMPORTANT
We suggest replacing the liner currently installed into the injector with
the HS/SPME straight glass empty liner PN 453A1335.
CAUTION
Be careful not to break the glass liner when removing it.
Glass splinters might fall into the lower part of the vaporization chamber.
Septum Cap
Ring nut
Septum holder/Liner cap with septum