Measuring the FID Gas Flows
Use this procedure for manually measuring the FID fuel and make-up gases.
To manually measure the FID flow rates
1. Put the GC in standby condition.
2. Turn the flame off. The fuel gases, hydrogen, and air, are automatically closed.
3. Cool the detector to room temperature.
4. Turn the make-up gas off.
5. Power off the GC.
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC, and
from the wall outlet.
6. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet
7. Open the module flap cover.
8. Disconnect the signal, glow-plug, and polarizing cables from their contacts on the cell top
cover. See
Materials needed
Forceps or tweezers
T10 Torxhead screwdriver
Measuring Tool - Flowmeter Adapter
Electronic flowmeter (Thermo Scientific GFM Pro Flowmeter, or equivalent)
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.