3
Performing Injectors Routine Maintenance
Maintaining a Split/Splitless Injector with Backflush (SSLBKF)
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
101
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
3. Turn the carrier gas off, and wait for the carrier pressure to go to zero.
4. Put the autosampler away if present.
5. Open the module flap cover.
6. Remove the top parts of the injector.
a.
Unscrew the septum cap of the injector.
b. Unscrew the ring nut.
c.
Remove the septum holder/liner cap with septum from the injector body head.
7. Replace the head body o-rings.
a.
Use tweezers to remove the body head internal and external o-rings, and replace them
with new o-rings.
8. Reinstall the top parts of the injector.
a.
Place the septum holder/liner cap with the septum on the body head of the injector
and fix it by screwing the ring nut.
b. Screw and tighten the injector cap to finger-tight.
9. Close the module flap cover.
10. If present, move the autosampler towards the module to restore the original alignment.
11. Turn the carrier gas on.
12. Set the normal injector, detector, and GC working conditions.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.
CAUTION
Do not overtighten the septum cap because you might damage the septum and
affect instrument performance.