7
Detectors Advanced Maintenance
Replacing the FPD Photomultiplier Tube
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
321
9. Remount the photomultiplier tube assemble.
a.
Reinsert the photomultiplier tube housing into the photomultiplier tube holder.
Holding the photomultiplier tube holder with one hand, use the other hand to
reinsert the photomultiplier tube housing into the holder. During the insertion meets
some resistance due to the presence of an o-ring inside the holder.
See
Figure 320.
Photomultiplier Tube Housing Reinstallation (1)
b. Fix the photomultiplier tube housing to the photomultiplier tube holder.
Carefully rotate the photomultiplier tube housing up to matching its fixing holes to
the fixing holes on the photomultiplier tube holder. See
.
Figure 321.
Photomultiplier Tube Housing Reinstallation (2)
c.
Insert the two fixing screws into the corresponding fixing holes, then tighten the
screws using a crosshead screwdriver.