8
Installing Optional Kits
Installing a FTIR Make-up Module
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
397
The standard operating value is in the range 0.2- 0.3 mL/min. Refer to the
Nicolet iS50 FT-IR
Spectrometer Manual
for details.
The components of the FTIR module are shown in
Figure 405.
FTIR Make-up Module
To add a FTIR Make-up module
1. Put the GC in standby condition.
2. Cool the oven, injector, and detector to room temperature.
3. Close the gas supplies.
4. Power off the GC.
ATTENTION
The flow of the carrier gas depends on the inlet pressure. Set the carrier gas
flow according to the working pressure.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.
Bottom Fitting
Nut and Ferrules
Flow Regulation Screw
Open/Close Knob
Needle Valve VICI-CNV1A10S1
ON/OFF Knob:
Turn clockwise for closing
Turn counter-clockwise for opening
Flow Regulation Screw:
Turn clockwise for decreasing
Turn counter-clockwise for increasing