8
Installing Optional Kits
Installing the Hot Injection Adapter Kit on the SSL/SSLBKF Injector
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
401
To install the hot injection adapter on the injector
1. Assemble the hot injection adapter if not already done. See
.
Figure 409.
Adapter Assembly
a.
Place the liner seal on the liner.
b. Insert the liner into the stem.
c.
Couple the head and the stem paying attention to properly align the corresponding
holes doing match the polarizing guides on the head and on the flange of the stem.
d. Fix the stem to the head screwing the three fixing screws using a T8 Torxhead
screwdriver.
2. Put the GC in standby condition.
3. Cool the oven, injector, and detector to room temperature.
4. Close the gas supplies.
5. Power off the GC.
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
6. Remove the column end from the injector.
Note
The stem conducts heat from the body of the SSL/SSLBKF injector to the
head of the adapter allowing a hot injection.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.
Liner Seal
Liner
Stem
Fixing
Screws
Guides