9
Adding Modules
Adding a NPD Detector Module
436
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 450.
NPD Thermionic Source Power Module Connections
1. AC input connector.
The module has a power rating of 120/230 Vac; 50/60 Hz; 65VA.
2. A 5-pin connector marked
NPD Module
(
2
)
for the connection of the signal cable
coming from the NPD detector module.
3. A 2-pin connector marked
NPD Source
(
3
)
for the connection of the thermionic source
assembly cable coming from the NPD detector module.
One protecting fuse is present inside the module. See
To add a new NPD detector module
1. Put the GC in standby condition.
2. Cool the oven, injector, and detector to room temperature.
3. Close the gas supplies.
4. Power off the GC.
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
6. Remove the dummy module from the position where the detector module must be
installed.
a.
Open the module flap cover.
Table 18.
Aux Temperature/Cryo Module Protecting Fuses
Fuse
Type
Protections
F1
F4A 250V; (5 x 20 mm)
Circuit for the generation of the current for the thermionic source
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.
2
3
1
Содержание TRACE 1300
Страница 1: ...TRACE 1300 and TRACE 1310 Gas Chromatographs Hardware Manual PN 31715002 Revision F January 2016 ...
Страница 4: ......
Страница 16: ......
Страница 94: ......
Страница 224: ......
Страница 450: ......
Страница 578: ......
Страница 610: ......
Страница 620: ......
Страница 624: ......