Performing Detectors Routine Maintenance
Maintaining an Electron Capture Detector (ECD)
170
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 152.
ECD Cables Disconnection
a.
Unscrew and pull out the straight plug crimp connector of the signal cable from the
collecting electrode (anode) bulkhead jack.
b. Pull out the quick coupling straight jack connector of the excitation cable from the
polarizing electrode bulkhead jack.
c.
Carefully move the cables in order to have free space for handling the detector.
8. Remove and clean the collecting electrode (anode).
a.
Unscrew the screw plug with its seal hiding the collecting electrode (anode) using the
T20 Torxhead screwdriver.
b. Using a little flathead screwdriver, unscrew and remove the collecting electrode
(anode). See
.
Signal Cable
Excitation Cable