1
Installation
Coupling to a Mass Spectrometer
30
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 38.
Left/Right Exterior Oven Wall Box View
c.
On the interior of the oven box remove the partial cut plate from the corresponding
duct. See
Figure 39.
Left/Right Interior Oven Wall Box View
d. Place the left/right panel for MS and attach the screw holding it in place.
3. Introduce the transfer line inner tube into the oven through the duct provided.
4. Attach the transfer line to the GC column using the proper nut and ferrule.
5. Close the front door of the GC.
6. Connect the transfer line heater if required.
a.
Connect the transfer line heater to the
Aux Temperature/Cryo Module
. See
Right Oven Wall
Left Oven Wall
IMPORTANT
Save the pre-shaped plug of insulating material in a safe place because it
could be reused.
Left Oven Wall
Right Oven Wall
CAUTION - INSTRUMENT DAMAGE:
Make sure that the GC column has been
conditioned before connecting it to the transfer line. The material released from the
column (column bleed) during conditioning may contaminate the detector.