7
Detectors Advanced Maintenance
Cleaning or Replacing the FPD Filter-side Heat Shields
302
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
7. Remove the photomultiplier assembly and the filter.
a.
Loosen the knurled nut that fixes the photomultiplier assembly and remove it from
the detector body. See
Figure 288.
Photomultiplier Assembly Removal
b. Remove the interferential filter from its housing, handling it very gently. Keep it
using a clean paper towel. See
Figure 289.
Filter Removal
Knurled Nut
CAUTION
The photomultiplier tube could damage if exposed to ambient light with the
excitation voltage
On
. Make sure the power supply has been switched off before
disconnecting the tube from the detector body.
CAUTION
Filters are fragile. Pay attention not to let the filter fall down and damage.