Detectors Advanced Maintenance
Cleaning or Replacing the FPD Flame-side Heat Shields
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
315
Figure 311.
Photomultiplier Assembly Remounting
12. Reinstall the FPD detector on the base.
a.
Place the detector on its base body, ensuring that the aluminium ring has been
inserted in the correct position, then tighten the fixing nut. See
Figure 312.
FPD Detector Reinstallation
b. Reconnect the signal, excitation voltage and ignition/heating cables to the detector.
See