10
Adding Systems
Adding the Hydrogen Sensor
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
547
11. Connect the sensor cable to the 10-pin connector marked
J8 Hydrogen Sensor
of the
Backplane board.
12. Close the oven door.
13. Reinstall the back cover.
a.
Reconnect the ground wire to the back cover terminal.
b. Replace the cover proceeding in the reverse order in which it was removed.
14. Reinstall the left side panel.
a.
Plug the ground wire to the panel.
b. Reinstall the panel proceeding in the reverse order in which it was removed.
15. If external modules are present, plug the power cable to the AC Input connector of each
external module, and to the wall outlet.
16. Power on the GC.
a.
Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker), located at the back of the instrument, to the
position I.
17. Set the normal injector, detector and GC working conditions.