8
Installing Optional Kits
Installing a FTIR Make-up Module
396
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Installing a FTIR Make-up Module
This section provides instructions for installing the FTIR Make-up module on your
TRACE 1300/1310 for the connection with the Thermo Scientific Nicolet iS50 FT-IR
Spectrometer
.
According to the configuration of your TRACE 1300/TRACE 1310, a dummy
module is present into the free site where the detector module is not installed.
See
Figure 404.
Add a Front/Back Detector Module
The FTIR Make-up module must be installed into the detectors housing, and supplied
through the line of the front/back carrier gas.
The module is equipped with an ON/OFF needle valve for the regulation of the gas flow.
• Open the needle valve manually turning the knob counter-clockwise.
• Close the needle valve manually turning the knob clockwise.
Perform the regulation of the gas flow using a little flat head screwdriver.
• Increase the gas flow turning the flow regulation screw counter-clockwise.
• Decrease the gas flow turning the flow regulation screw clockwise.
Dummy Back Detector Module
Dummy Front Detector Module
ATTENTION
Where a dummy module is installed, the gas connection is blocked by a
plug.
IMPORTANT
If the front/back carrier gas is not already used by an injector module, you
must install the front/back carrier gas inlet, and connect the carrier gas line as described in
the section
“Making the Gas Supply Plumbing Connections”
.