Installing Optional Kits
Installing the Packed Column Adapters
338
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Installing the Packed Column Adapters
This section provides instructions for installing the adapters for connecting 1/8 or 1/16-in.
OD metal packed columns to the bottom of the SSL injector, FID, NPD, ECD, TCD, and
FPD. For the connection to a PDD detector see the
Caution
note below.
Introduction
The adapters are provided with the kit PN 19050758. The kit includes the following parts:
Getting Started
Install the column adapters following the instructions in the next operating procedures.
•
“To install the packed column adapters”
•
“To connect a new packed column to the injector and detector adapters”
WARNING
Do not try to connect packed columns if the lengths of the column ends
above the end of the ferrule are not compatible with the adapters. In this case we suggest
to connect a new packed column by using the nuts and ferrules provided with the kit.
CAUTION
Connecting a packed column to a PDD detector
: An adapter to connect the
PDD packed column adapter can be assembled using a piece of tube, the 1/16-in. nut and
gold plated ferrule, and a 1/8-in. to 1/16-in. reducing union in the standard outfit.
Table 14.
Packed Column Adapters and 1/2-in. Wrench
Part
Description
Adapter for the connection of the 1/8-in. OD packed column
to the SSL injector
Adapter for the connection of the 1/16-in. OD packed
column to the SSL injector
Adapter for the connection of the 1/8-in. OD packed column
to the FID, NPD, ECD, TCD, or FPD detector
Adapter for the connection of the 1/16-in. OD packed
column to the FID, NPD, ECD, or FPD detector
1/2-in. wrench