Installing Optional Kits
Installing the NoVent Microfluidics
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
377
To connect the NoVent Microfluidics module to the TRACE 1300/1310 GC
1. Cool the oven, injector or injectors, transfer line, ion source, and any installed GC
detectors to room temperature and shut down the GC.
2. Push down on the power switch to power off the GC.
3. Shut down the TSQ 8000 Series or ISQ Series mass spectrometer using the software. See
your mass spectrometer's user guide for more information. The heaters and the
turbo-molecular pump power off.
4. If you are using hydrogen as a carrier gas, unscrew the hydrogen safety screw on the front
door of the mass spectrometer.
Figure 373.
Finding the Vent Valve on the Mass Spectrometer
ATTENTION
Where a dummy module is installed, the gas connection is blocked by a
plug.
WARNING - BURN HAZARD:
The injectors, detectors, oven, and transfer line may be hot.
Allow them to cool to room temperature before touching them.
WARNING - FIRE HAZARD:
If you are using hydrogen, do NOT reach over the top of the
instrument to power it off. Instead, reach around the right side or go to the back of the
instrument and flip down the power switch.
Vacuum Interlock Shield
Power Switch
Vent Valve Knob