Adding Modules
Adding a Generic Detector Interface
488
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 520.
Gas Lines Connection to the GDI Electrical Interface
Connecting Heater and Signal Cables
If required by the third-party detector, connect the heater cable, the signal cable, or both.
To connect the heater and signal cables
1. Using the cable provided connect the 5-pin connector marked
Heater
on the GDI
electrical interface to the detector connector on the GDI mechanical module.
See
Figure 521.
Heater and Signal Connectors
Gas 3
Gas 2
Gas 1
Gas 3
Gas 2
Gas 1
IMPORTANT
The maximum nominal inlet pressure for all the inputs is 1000 kPa
(145 psig). The working inlet pressure range is from 400 kPa (58 psig) to 1000 kPa (
145 psig).
From Third-party
Detector
To Heater
Connector
on Detector
Module