Adding Modules
Adding a Generic Detector Interface
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
479
c.
Finger-tighten the hexagonal screws slightly, or use a 10-mm wrench.
3. Connect the 26-pin connector of the detector cable coming from the GDI electrical
interface to the backplane board in the electronic compartment.
a.
On the backplane board disconnect the flat cable from the connector marked
J13
DET.FRONT
if the GDI mechanical module is installed in the
Front
site, or from
the connector marked
J5 DET.REAR
if the GDI mechanical module is installed in
the
Back
site. See
Figure 507.
GC Backplane Board Layout
b. Connect the detector cable to the connector
J13 DET.FRONT
, or
J5 DET.BACK
accordingly.
Note
Always keep the hexagonal screws in their place. This allows you an easier
removal of the GDI electrical interface when necessary.
ATTENTION
If an GDI mechanical module will be installed on the TRACE
1300/TRACE 1310 GC, continue the procedure from the
. If an GDI mechanical
module will be installed on the TRACE 1310 Auxiliary Oven, jump to the section
“Connecting a GDI Interface to the TRACE 1310 Auxiliary Oven”
J13 DET.FRONT Connector
J5 DET.REAR Connector