Installing Optional Kits
Installing the NoVent Microfluidics
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
387
10. While holding the column measuring tool steady, tighten the transfer line nut with a
1/4 in. wrench until the column just stops moving in the ferrule.
11. Turn the transfer line nut 1 flat backward so the column is able to move in the ferrule
with slight resistance.
12. Line up the outlet of the column with the arrows on the end of the column measuring
tool.
Figure 391.
Lining Up the Column in the Column Measuring Tool
13. Place a septum with a notch cut into it behind the transfer line nut. The septum marks
the place on the column where it should exit the nut.
Figure 392.
Positioning the Septum
14. Pull the column back from the transfer line nut. Do not move the septum from its
position on the column.
Column Outlet
Transfer Line
Nut
Septum
Column
Measuring Tool