4
Performing Detectors Routine Maintenance
Maintaining a Flame Photometric Detector (FPD)
188
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
5. If other external modules are present, unplug the power cable from the AC Input
connector of each module, and from the wall outlet.
6. Remove the FPD detector from the base.
a.
Disconnect the signal, excitation voltage, and ignition/heating cables from the
detector. See
.
Figure 174.
FPD Cables Disconnection
b. Using the tool provided with the system, loosen the fixing nut on the base of the
.
Figure 175.
FPD Detector Removal
FPD Fixing Nut
Aluminium Ring