3
Performing Injectors Routine Maintenance
Maintaining a Programmable Temperature Vaporizing Injector with Backflush (PTVBKF)
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
133
13. If present, move the autosampler towards the module to restore the original alignment.
14. Turn the carrier gas on.
15. Set the normal injector, detector, and GC working conditions.
Replacing the PTVBKF Broken Liner
To replace a broken liner
Figure 112.
PTVBKF Injector: Broken Liner Replacement
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
3. Turn the carrier gas off, and wait for the carrier pressure to go to zero.
Septum
Septum Cap
Liner Cap
Liner Seal
Liner
Silver Seal
Terminal Fitting
Bottom Fittings
Split Nut
Ferrule
Backflush Line