8
Installing Optional Kits
Performing the Dual FPD Detector Configuration
414
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
c.
Insert the handle of a screwdriver or other un-sharpened tool in the combustion
chamber and push the flame-side heat shield and its graphite seal out from its
housing. Act gently to avoid breaking of the heat shield. See
.
Figure 427.
Filter-side Heat Shield Removal
d. Remove with care traces of graphite due to the breaking of the seal.
e.
Reinstall the mirror plug.
Assembling the primary FPD detector in Dual FPD configuration
Perform the Dual FPD configuration on the primary FPD detector using the components
removed from the secondary FPD detector.
9. Remove the mirror plug from the cell of the primary FPD detector.
a.
Loosen the Allen screws that fix the mirror plug and remove it. See
.
Figure 428.
Mirror Plug Removal
10. Insert the filter support and the flange assembly into the primary FPD detector.
a.
Insert the heat shield into its housing inside the detector body.
b. Insert the spacer into the flange, letting it stand out for about 5 mm.
CAUTION
While pushing out the heat shield, pay attention not to damage the ignition
coil.
Содержание TRACE 1300