5
GC Main Frame Advanced Maintenance
Replacing the Oven Heater Baffle
212
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
To replace the oven heater baffle
1. Put the GC in standby condition.
2. Cool the oven, injectors and detectors to room temperature.
3. Power off the GC.
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
4. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
5. Remove the back cover.
a.
Using a T20 Torxhead screwdriver, remove the four screws that secure the back cover
to the GC. See
.
Figure 201.
Back Cover Removal
b. Lift the cover off using the cover handle. Be aware that the ground wire is attached to
the back panel.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.
Note
Pay attention to the positioning of the ground wire plug, so it can be
reconnected in the same way it was removed.
Back Cover Fixing Screws