background image

Pipetman

®

continuously adjustable 
digital microliter pipettes

8 pipettes for volume
ranges from 
0.1 µL to 10 mL

P-2

P-10

P-20

P-100

P-200

P-1000

P-5000

P-10ML

Pipetman P-200 shown

Содержание P-10

Страница 1: ...Pipetman continuously adjustable digital microliter pipettes 8 pipettes for volume ranges from 0 1 L to 10 mL P 2 P 10 P 20 P 100 P 200 P 1000 P 5000 P 10ML Pipetman P 200 shown...

Страница 2: ...and Repair 10 Contacting RAININ 10 Performance Specifications 11 Replacement Parts Back Cover FIGURES Figure 1 Pipetman P 200 1 Figure 2 Operating Pipetman 4 Figure 3 Removing Tip Ejector Arm 9 DIAGR...

Страница 3: ...which reads in microliters The digital volume indicator simplifies volume setting and virtually elimi nates calculation errors All Pipetman models have a highly polished stainless steel piston patente...

Страница 4: ...rements for each Pipetman model are shown below Pipetman Range L Smallest Model Adjustable Recommended Increment L P 2 0 to 2 0 1 to 2 0 002 P 10 0 to 10 0 5 to 10 0 02 P 20 0 to 20 2 to 20 0 02 P 100...

Страница 5: ...he desired setting turn the dial 1 3 revolution higher than desired and reset the volume The friction ring prevents unintentional volume changes 3 Attach a new disposable tip to the pipette shaft Pres...

Страница 6: ...p Selection I Tips must seal properly on the shaft to assure an air tight seal and avoid leaks or poor accuracy I Tips must be soft and flexible so that the shaft is not scratched or worn prematurely...

Страница 7: ...mm P 200 50 200 L 3 6 mm P 1000 1000 1000 L 3 6 mm P 5000 500 5000 L 6 10 mm P 10ML 1 mL 10 mL 6 10 mm Tip immersion depth is important If exceeded the volume measured will be inaccurate possibly out...

Страница 8: ...e any excess liquid from the outside of the tip without touching the orifice 5 To dispense rest the end of the tip against the vessel wall and press the plunger to the first stop Hold this position a...

Страница 9: ...ntact with corrosive fumes may corrode the piston This will result in premature seal wear and may require refinishing or replacement of the piston Exposure of internal Pipetman com ponents to corrosiv...

Страница 10: ...803D which use a grease seal Leaks Inaccurate Sampling 1 Loosened shaft Tighten coupling by hand 2 Split or cracked shaft Remove the tip ejector and inspect the shaft for fracture or split end Replace...

Страница 11: ...ector opening over the end of the Pipetman shaft 5 Push the plastic collar into the handle until the tip ejec tor is securely in place Make sure that tips will fit prop erly on the shaft Figure 3 Tip...

Страница 12: ...Tel 858 320 0446 It is NOT necessary to recalibrate the pipette after changing the seal or shaft Recalibration of the pipette is only necessary when the piston is replaced and should only be done by q...

Страница 13: ...r both the density of water and evaporation where applicable Consult the Rainin publication Procedure for Evaluating Pipette Accuracy and Precision AB 15 for further information or download a copy fro...

Страница 14: ...e diagrams are for dual adjustment Pipetman and cannot be used on older Pipetman models For convenience old style part numbers are also listed on the back page 12 P 2 P 10 P 20 P 100 A B M L K J H G F...

Страница 15: ...13 P 200 P 1000 P 5000 P 10ML A B M L K J H G F C D E A B G F C D E A B D N N A B F G D G F...

Страница 16: ...a n a n a n a 6190 164 6190 164 Filters 1000 n a n a n a n a n a n a 6190 165 6190 165 This column is for P 10ML after S N N05806D For P 10ML pre S N N05803D same as listed above except Plunger rod 23...

Отзывы: