8
Installing Optional Kits
Installing the Packed Column Adapters
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
341
8. Install the adapter on the bottom the detector interposing the washer. See
.
Figure 342.
Adapter Installation on the Bottom of the Detector
a.
Finger-tighten the retaining nut of the adapter until it start to grip the bottom of the
detector.
b. Use the 1/2 wrench to firmly tighten the retaining nut of the adapter with the base
seal from the bottom of the injector.
To connect a new packed column to the injector and detector adapters
Before you begin, verify that the proper adapters are installed on the injector and detector
side.
1. Preparing the metal packed column.
a.
Using a 1/4-5/16-in. wrench, remove the column nut and ferrules from the base of
the injector adapter.
b. Using a 1/4-5/16-in. wrench, remove the column nut and ferrules from the base of
the detector adapter.
c.
Slide the column nut and the ferrules onto the packed column injector and detector
ends in the order and direction as shown in the
.
Adapter for 1/8-in. OD Packed Column
Adapter for 1/16-in. OD Packed Column
Retaining Nut
Washer